Please use this identifier to cite or link to this item: http://library.iigm.res.in:8080/xmlui/handle/123456789/1217
Full metadata record
DC FieldValueLanguage
dc.contributor.authorMahavarkar, Prasanna
dc.contributor.authorJohn, Jacob
dc.contributor.authorDhapre, V.Y.
dc.contributor.authorDongre, Varun
dc.contributor.authorLabde, Sachin
dc.date.accessioned2011-06-18T02:52:19Z
dc.date.accessioned2021-02-12T09:22:16Z-
dc.date.available2011-06-18T02:52:19Z
dc.date.available2021-02-12T09:22:16Z-
dc.date.issued2018
dc.identifier.citationGeoscientific instrumentation methods and data systems, 7, 143–149, doi: 10.5194/gi-7-143-2018en_US
dc.identifier.urihttp://localhost:8080/xmlui/handle/123456789/1217-
dc.description.abstractA tri-axial square Helmholtz coil system for the study of palaeomagnetic studies, manufactured by GEOFYZIKA (former Czechoslovakia), was successfully commissioned at the Alibag Magnetic Observatory (IAGA code: ABG) in the year 1985. This system was used for a few years, after which the system encountered technical problems with the control unit. Rectification of the unit could not be undertaken, as the information document related to this system was not available, and as a result the system had been lying in an unused state for a long time, until 2015, when the system was recommissioned and upgraded to a test facility for calibrating the magnetometer sensors. We have upgraded the system with a constant current source and a data-logging unit. Both of these units have been designed and developed in the institute laboratory. Also, re-measurements of the existing system have been made thoroughly. The upgraded system is semi-automatic, enabling non-specialists to operate it after a brief period of instruction. This facility is now widely used at the parent institute and external institutions to calibrate magnetometers and it also serves as a national facility. Here the design of this system with the calibration results for the space-borne fluxgate magnetometers is presented.en_US
dc.language.isoen_USen_US
dc.subjectAlibag magnetic observatoryen_US
dc.subjectTri-axial square Helmholtz coil systemen_US
dc.subjectPaleomagnetic studiesen_US
dc.subjectFluxgate magnetometeren_US
dc.titleTri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facilityen_US
dc.typeArticleen_US